Pulse homogenizer use precautions and cleaning and maintenance

Pulse homogenizer use precautions

1. In order to effectively reduce the vibration during the operation of the equipment, the internal circuit board is adversely affected. The supporting feet of the instrument are made of soft plastic anti-vibration rubber feet. Pay attention to lifting the equipment to avoid dragging.

2.
Pulse homogenizer is mainly used in laboratory room, the use temperature can not be lower than 10 ° C;

3. If the temperature difference between the storage or transportation temperature of the instrument and the ambient temperature of the operating equipment is ≥10 °C, leave it at least at room temperature for two hours before use to avoid damage to the equipment caused by condensation caused by temperature difference inside the equipment.

4. If the conditions of use are outside the scope of the manual, it may cause damage to the instrument;

5. The maximum processing time of the sample processor can be set to 11 minutes and 59 seconds. If the instrument is continuously used continuously, the temperature of the generator is higher than the safe use temperature. At this time, the internal automatic detection mechanism of the instrument will force the motor to stop working. In order to protect the safe operation of the motor.

6. Before connecting the power supply, please check if the power supply specifications are suitable (if you have any doubts, consult a professional electrician). Check that the ventilation slots are clear before use.

Pulse homogenizer cleaning and maintenance

The pulsed homogenizer design ensures that the spilled sample processing fluid remains in the processor's processing chamber, ensuring that the power is removed during the cleaning process.
When cleaning the external surface of the processor and the oscillator, clean it with mild soapy water and then dry it gently with a soft cloth (if the inside of the processor is contaminated with bacteria, use 70% isopropyl alcohol solution) .
Do not use acid, alkali, and chlorine and salt solutions to clean the outside of the processor, and avoid organic solvents from damaging the processor.
Do not use scouring or scraping during the cleaning process, otherwise it will cause irreversible damage to the processor.
If the cleaning method used is different from the one mentioned above, consult the manufacturer or agent to see if the method used will damage the instrument.

General note:
The microswitch on the sample processor processing chamber ensures that the instrument will not operate when the injection door is not properly closed. Do not perform any behavior that prevents the microswitch from functioning properly. When the running noise is increased or changed, report it to the manufacturer or supplier in time so that it can be disposed of in a timely manner.